Baxter TruSystem Operating Tables Can Drop Unexpectedly During Use

3 minutes ago
2 min read

We spend time getting a patient positioned, securing lines and giving anesthesia access to the airway. Now imagine the OR table suddenly drops to its lowest position and won’t raise again. Even if the surgical team stops immediately, everyone in the room has something to account for.

On November 3, the FDA posted Class II recall records for two Baxter TruSystem operating table corrections that began in September. Both involve unintended downward movement, but they cover different issues and have different instructions. The model and serial number tell you which notice applies.


The September 14 correction involves belt tension

The FDA record for the TruSystem 7000 U Operating Table (dV) lists serial numbers 110188921, 110188922 and 110206662. Baxter identified units that had undergone spindle replacement without the required belt-tension verification. If the belt tension is outside specification, the table column could move downward unexpectedly.

For this correction, Baxter said customers could continue using the affected tables while the company schedules the belt-tension verification. Facilities should use the full notice and their Baxter correspondence to confirm the action for each unit.


The September 29 correction involves wear in the main lift

The separate FDA record for the TruSystem 7500, catalog number 4091000, describes unexpected wear in the main lift assembly. Affected tables may drop to the lowest position and may be unable to rise again. The FDA record includes the affected serial numbers.

While waiting for the correction, Baxter directs customers to monitor the table during use, following the applicable instructions for use. If the affected table makes an unusual noise, discontinue use and contact Baxter Technical Service. Baxter also asks facilities to post the notice near affected tables until the correction is complete.


Why the whole room needs to know

An abrupt drop can displace a patient, compromise the airway, cause injury or interrupt the procedure. Biomed can match a serial number and coordinate service, but the people setting up and using the room also need to know which notice applies and what it says. That matters when a table is moved between rooms or when a different team comes in for the next case.

If your room were assigned a different TruSystem table tomorrow, how would your team confirm whether it has an open correction?


U.S. Food and Drug Administration. “Class 2 Device Recall TruSystem 7000 U Operating Table (dV).” FDA Medical Device Recall Database, 3 Nov. 2026, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=223130.

U.S. Food and Drug Administration. “Class 2 Device Recall TruSystem 7500.” FDA Medical Device Recall Database, 3 Nov. 2026, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=223352

 
 
 

Recent Posts

See All

Comments


For the love of surgery surgical tech

Empowering surgical professionals through education, connection, and real-world experience.

  • Instagram
  • TikTok
  • YouTube
  • LinkedIn
  • Facebook

Join Our Newsletter

Stay inspired and up to date on new workshops and resources by becoming a member & joining our email list.

© 2026 For The Love of Surgery. All Rights Reserved.

For the Love of Surgery is a 501(c)(3) organization and

your donation is tax deductible within the guidelines of US law.

Tax ID: 41-3695452​

4651 Mallow St, Houston, TX 77051
| Privacy Policy | Terms of Use | Disclaimer |

bottom of page